2.4mm Va-Lcp Intercarpal Fusion System
510(k) K103243
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK103243 | Date21 Jan 2011 | StatusActive |
Recalls
-
- Terminated Z-0880-2020
- Terminated Z-0881-2020
- Terminated Z-0882-2020
- Terminated Z-0879-2020
- Terminated Z-0883-2020
There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.
-
- Terminated Z-2389-2016
It was reported that the specified lots of the 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm can break between the head and the shaft. This resulted from the neck feature being too close to the StarDrive feature leaving a thin wall between neck and drive.
-
- Terminated Z-0323-2016
It was discovered that the above part number and lot of 2.4mm Locking Screws may contain the incorrect package insert (GP2613, VA-LCP Distal Radius System). The correct package insert for the above part is GP2615 (Modular Mini Fragment LCP System).
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- HRS
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3030
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent