Skip to content

3DQUORUM Software

by Hologic

3DQUORUM Software (SW-05673) is a stationary mammographic x-ray system (FDA Class III; FDA product code OTE, Digital breast tomosynthesis) made by Hologic, approved under FDA PMA P080003 on 11 Feb 2011 and listed in FDA GUDID since 18 Sep 2019 (UDI-DI 15420045512313), active as of 23 Oct 2026. 3 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

PMA

Valid until

No expiry

PMA approval is not time-limited

FDA recalls

3 ongoing

This device is named in the recall

Documents

PMA P080003

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number15420045512313 UDI-DI Date18 Sep 2019ListedStatusListed GUDID: in commercial distribution
United StatesFDA PMAAPPR from GUDID Approved NumberP080003 Date11 Feb 2011DecidedStatusActive

Official records

From the FDA record

Device description
3DQuorum® is an image processing software application for post-processing tomosynthesis data captured on a Hologic tomosynthesis imaging system, into a combined 3D image.
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 23 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA PMA P080003 decided 11 Feb 2011 Active
  • FDA GUDID UDI-DI 15420045512313 listed 18 Sep 2019 Listed

Recalls

4 recalls · 3 ongoing · FDA

    • Ongoing Z-2168-2026 FDA Class II initiated 8 Apr 2026

    Complaints have been received of systems developing loose, missing, or broken internal bolts over time.

    • Ongoing Z-3248-2024 FDA Class II initiated 19 Aug 2024

    Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may appear blurry and result in misdiagnosis with the possibility being a false negative diagnosis

    • Ongoing Z-1144-2023 FDA Class II initiated 16 Jan 2023

    C-Arm unexpected movement may cause blunt trauma should the tube arm impinge upon an individual

    • Terminated Z-1343-2020 FDA Class II initiated 21 Jan 2020 on shared PMA P080003

    The recalling firm has received reports from the field of the C-arm on the Selenia Dimensions system seizing or dropping during field repairs and cleaning of the systems.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

6 dated events

  1. Approved FDA PMA P080003
  2. Listed FDA GUDID 15420045512313
  3. Recall initiated FDA recalls Z-1343-2020
  4. Recall initiated FDA recalls Z-1144-2023
  5. Recall initiated FDA recalls Z-3248-2024
  6. Recall initiated FDA recalls Z-2168-2026

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA PMA Active

Specification

GUDID

Trade name
3DQuorum
Generic name
Stationary mammographic x-ray system, digital
Model / Ref / Catalog
3DQUORUM SOFTWARE, SW-05673
Primary DI / UDI-DI
15420045512313
PMA number
FDA product code
OTE
GMDN term
Stationary mammographic x-ray system, digital
Device class
Class III
GUDID distribution status
In Commercial Distribution

Similar devices in this category

Other devices filed under Stationary mammographic x-ray system, digital.

Other devices from Hologic

See all devices from Hologic →

Properties

  • Prescription use
  • Lot / batch number
Other properties (13, all answered No)