Skip to content

4D Integrated Treatment Console

by Varian Medical Systems 2 variants

510(k) K091132

4D Integrated Treatment Console is a medical device (FDA Class II; FDA product code IYE, Accelerator, Linear, Medical) made by Varian Medical Systems, cleared under FDA 510(k) K091132 on 17 Jun 2009, active as of 22 Oct 2026. 5 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

5 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK091132 Date17 Jun 2009ClearedStatusActive

Official records

From the FDA record

Received by the FDA
20 Apr 2009
FDA decision
17 Jun 2009 (Substantially Equivalent)
FDA review time
58 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K091132 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

5 recalls · FDA

    • Terminated Z-0691-2012 FDA Unclassified (pre-2012) initiated 12 Dec 2011

    Varian has discovered that on 4D Integrated Treatment Console and TrueBeam, the ad-hoc CIAO(Complete Irradiated Area Outline) calculation results in the CIAO generation that is different from the CIAO in RT Chart. The 4DITC algorithm adds dose for the control point at field edge, which should not be added according to the system specifications.

    • Terminated Z-0715-2012 FDA Unclassified (pre-2012) initiated 12 Dec 2011

    When multiple conical collimator treatment plans are scheduled to be treated in a session, the Plan Label should be displayed on the BCCV dialog box should match the Plan Label on the 4DITC. However, the name of the plan is not being updated correctly.

    • Terminated Z-3194-2011 FDA Unclassified (pre-2012) initiated 21 Jul 2011

    The 4D Integrated Treatment Console may erratically fail to load the planned physical wedge accessory without warning or notification when receiving an approved plan from the oncology information system.

    • Terminated Z-2119-2011 FDA Unclassified (pre-2012) initiated 28 Feb 2011

    A software anomaly, in that the Multileaf Collimator (MLC) message displayed on the 4DITC on versions 8.6 and below, does NOT interlock beam delivery.

    • Terminated Z-0891-2011 FDA Unclassified (pre-2012) initiated 16 Oct 2009

    Software anomaly may result in a design deficiency causing Plan Edit Settings to malfunction under specific circumstances, where the user selects to use a Cone Beam CT set up field.

Timeline

6 dated events

  1. Cleared FDA 510(k) K091132
  2. Recall initiated FDA recalls Z-0891-2011
  3. Recall initiated FDA recalls Z-2119-2011
  4. Recall initiated FDA recalls Z-3194-2011
  5. Recall initiated FDA recalls Z-0691-2012
  6. Recall initiated FDA recalls Z-0715-2012

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
IYE
Device class
Class II
Regulation
21 CFR Part 892.5050
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code IYE (Accelerator, Linear, Medical).

Other devices from Varian Medical Systems, Inc

See all devices from Varian Medical Systems, Inc →