  "format": "devicerts.device.v1",
  "cite_as": "4D Integrated Treatment Console, Varian Medical Systems. deviCERTS medical device catalogue. https://devicerts.com/devices/4d-integrated-treatment-console-iye (data as of 2026-11-23).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/4d-integrated-treatment-console-iye",
    "json_url": "https://devicerts.com/devices/4d-integrated-treatment-console-iye.json",
    "markdown_url": "https://devicerts.com/devices/4d-integrated-treatment-console-iye.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-23",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/4d-integrated-treatment-console-iye",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "4D Integrated Treatment Console",
    "registry_name": "4D INTEGRATED TREATMENT CONSOLE",
    "url": "https://devicerts.com/devices/4d-integrated-treatment-console-iye",
    "manufacturer": {
      "name": "Varian Medical Systems",
      "url": "https://devicerts.com/manufacturers/fda-fei-2916710",
      "country": null,
      "country_basis": null,
      "fda_registration_country": "United States"
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K091132",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "IYE"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K091132",
      "number_label": null,
      "device_udi_di": null,
      "date": "2009-06-17",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0691-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-12-12",
      "reason": "Varian has discovered that on 4D Integrated Treatment Console and TrueBeam, the ad-hoc CIAO(Complete Irradiated Area Outline) calculation results in the CIAO generation that is different from the CIAO in RT Chart. The 4DITC algorithm adds dose for the control point at field edge, which should  not be added according to the system specifications.",
      "applies": "this device",
      "via": "K091132"
    },
    {
      "number": "Z-0715-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-12-12",
      "reason": "When multiple conical collimator treatment plans are scheduled to be treated in a session, the\nPlan Label should be displayed on the BCCV dialog box should match the Plan Label on the\n4DITC. However, the name of the plan is not being updated correctly.",
      "applies": "this device",
      "via": "K091132"
    },
    {
      "number": "Z-3194-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-07-21",
      "reason": "The 4D Integrated Treatment Console may erratically fail to load the planned physical wedge accessory without warning or notification when receiving an approved plan from the oncology information system.",
      "applies": "this device",
      "via": "K091132"
    },
    {
      "number": "Z-2119-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-02-28",
      "reason": "A software anomaly, in that the Multileaf Collimator (MLC) message displayed on the 4DITC on versions 8.6 and below, does NOT interlock beam delivery.",
      "applies": "this device",
      "via": "K091132"
    },
    {
      "number": "Z-0891-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-10-16",
      "reason": "Software anomaly may result in a design deficiency causing Plan Edit Settings to malfunction under specific circumstances, where the user selects to use a Cone Beam CT set up field.",
      "applies": "this device",
      "via": "K091132"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K091132",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "IYE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.5050",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}