# 4D Integrated Treatment Console

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/4d-integrated-treatment-console-iye>
- Manufacturer: [Varian Medical Systems](https://devicerts.com/manufacturers/fda-fei-2916710) (FDA registration: United States)
- Name in the registry: 4D INTEGRATED TREATMENT CONSOLE
- Device class: Class II (US FDA)
- FDA product code: IYE

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K091132 · decision 2009-06-17 · decision SESE · status: Active

## Recalls

- Z-0691-2012 ·  · terminated · initiated 2011-12-12: Varian has discovered that on 4D Integrated Treatment Console and TrueBeam, the ad-hoc CIAO(Complete Irradiated Area Outline) calculation results in the CIAO generation that is different from the CIAO in RT Chart. The 4DITC algorithm adds dose for the control point at field edge, which should not be added according to the system specifications.
- Z-0715-2012 ·  · terminated · initiated 2011-12-12: When multiple conical collimator treatment plans are scheduled to be treated in a session, the Plan Label should be displayed on the BCCV dialog box should match the Plan Label on the 4DITC. However, the name of the plan is not being updated correctly.
- Z-3194-2011 ·  · terminated · initiated 2011-07-21: The 4D Integrated Treatment Console may erratically fail to load the planned physical wedge accessory without warning or notification when receiving an approved plan from the oncology information system.
- Z-2119-2011 ·  · terminated · initiated 2011-02-28: A software anomaly, in that the Multileaf Collimator (MLC) message displayed on the 4DITC on versions 8.6 and below, does NOT interlock beam delivery.
- Z-0891-2011 ·  · terminated · initiated 2009-10-16: Software anomaly may result in a design deficiency causing Plan Edit Settings to malfunction under specific circumstances, where the user selects to use a Cone Beam CT set up field.

## Identifiers

- 510(k) Number: K091132 [FDA 510(k)]
- FDA product code: IYE [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 892.5050 [FDA 510(k)]
- Advisory committee: Radiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: 4D Integrated Treatment Console, Varian Medical Systems. deviCERTS medical device catalogue. https://devicerts.com/devices/4d-integrated-treatment-console-iye (data as of 2026-11-23).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-23
- JSON: <https://devicerts.com/devices/4d-integrated-treatment-console-iye.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/4d-integrated-treatment-console-iye
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.