Skip to content

700-F Series Stimulation Electrodes

by Cardiotronics Systems, Inc.

510(k) K964469

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

4 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK964469 Date11 Dec 1996ClearedStatusActive

Recalls

4 recalls · FDA

  • 2 recalls, same reason

    • Terminated Z-0462-2015 FDA Class I initiated 6 Nov 2014
    • Terminated Z-0461-2015 FDA Class I initiated 6 Nov 2014

    Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.

    • Terminated Z-0856-2012 FDA Unclassified (pre-2012) initiated 31 Mar 2011 on shared 510(k) K964469

    The specific lot was assembled without the rivet cover.

    • Terminated Z-0735-2012 FDA Unclassified (pre-2012) initiated 9 Mar 2011

    The product lot was assembled with an R2 Connector, rather than a Zoll Connector.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LDD
Device Class
Class II
Regulation
21 CFR Part 870.5300
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent