8600 Vital Signs Monitor
510(k) K042569
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK042569 | Date29 Sep 2004 | StatusActive |
Recalls
-
- Terminated Z-0810-2014
The main board, P/N 301641, installed in some Medrad Veris MR Monitor units may be faulty and could lead to unexpected shutdown of the system while in use, resulting in the loss of vital signs information from the monitor.
-
- Terminated Z-1586-2013
The firm is recalling these power cables due to a latent design reliability issue and the potential for shorting which can result in heating/melting of the cable jacket.
-
- Terminated Z-0259-2007
Interference problems: certain invasive blood pressure transducers connected to these cables are susceptible to interference if exposed to static magnetic field strengths exceeding 300 gauss. In certain orientations, the transducer causes the invasive blood pressure readings on the physiological monitor to display inaccurate readings.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MWI
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.2300
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent