Abbott Architect B12
by Abbott Laboratories 510(k)
Active
510(k) K984108
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK984108 | Date3 Feb 1999 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0860-06
The 30-day onboard storage information is not included in the barcode labels for these two lots. As a result, the ARCHITECT system software is unable to track how long a reagent kit has been stored onboard the ARCHITECT instrument.
-
- Terminated Z-1452-04
Changes in the ARCHITECT B12 Reagent lots may adversely affect the shape of the calibration curve, over time. These changes may produce controls out of range low and similar decreases in patient sample values prior to current expiration of the reagents.
-
- Terminated Z-0435-05
Falsely elevated B-hCG results can occur when running B12 and B-hCG assays on the same analyzer due to carryover. B-12 results are not impacted.
Specification
- 510(k) Number
- FDA Product Code
- LIG
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1810
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent