Abbott Architect Testosterone
by Abbott Laboratories 510(k)
Active
510(k) K983212
Regulatory status
Recall ongoing on a related submission. 2 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK983212 | Date18 Nov 1998 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-0272-2022
- Ongoing Z-0271-2022
Twelve software-related issues affecting software version 9.41 and earlier were identified. The issues include bypassing error codes, incorrect calibration, loss of module communication leading to overfilling wash buffer containers, missing QC flags, signal spikes, abnormal optics reads, and incorrect sample aspiration. The issues may lead to inaccurate results with an unknown bias direction and magnitude for assays measuring cardiac, metabolic (including diabetes), toxicology, prenatal, therapeutic drug monitoring, infectious disease, oncology, hepatic, pancreatic, hematological, endocrinologic, inflammatory analytes.
-
- Terminated Z-1316-2010
ARCHITECT System Software v5.00 and v5.10 are incompatible with ARCHITECT LLS Board. Software patch re-establishes correct parameters.
-
- Terminated Z-0132-2010
Under certain conditions, the product's wash buffer containing sodium azide can come in contact with the wash zone mechanism ground cables on either of the two wash zones. The copper in the ground strap may corrode and form an unstable chemical substance (metal azide) that may be sensitive to direct pressure and impact.
-
- Terminated Z-0951-2007
- Terminated Z-0950-2007
One or more of four software issues : 1) calculation error for Architect cSystem assays using the Spline math model. 2) The system configuration option, Run controls onboard reagents by Kit, does not function correctly for assays requiring a standard sample dilution. 3) Infrequent data may be missing from automatically or manually printed reports generated on the Architect system. 4) Clinical Chem
-
- Terminated Z-0791-2007
- Terminated Z-0790-2007
If the ARCHITECT i2000/i2000sr System becomes contaminated with microbes generating folate-like by-products, the ARCHITECT Folate assay may experience calibration failures and/or shifts in results.
-
- Terminated Z-0415-2007
- Terminated Z-0416-2007
The re-designed ARCHITECT Temperature Controller Board has been identified as being susceptible to electromagnetic interference (EMI) in the laboratory. This can cause a board reset condition, stopping temperature control function of the board. Situation does not stop analyzer operation and does not generate an error condition alerting operator when condition occurs.
-
- Terminated Z-0818-04
When using a LIS the software system will send up a test result of up to 20 characters although the field is set to display only 15 characters per the RS-232 Manual.
-
- Terminated Z-0262-04
Sample Identification (SID) numbers of a bar coded sample may be truncated, or shortened, after being scanned by the bar code reader (e.g. an actual SID of 123456 is shortened to 1234)
-
- Terminated Z-0172-04
If the interlock cable connected to the door lid sensor is disconnected or fails, the lid is detected as closed even though it may be open.
-
- Terminated Z-1133-03
Under specific conditions, the system can allow an incorrect sample ID to be assigned to another sample in a different carrier.
Specification
- 510(k) Number
- FDA Product Code
- CDZ
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1680
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent