Abbott Plum a+infusion Pump, Model 11971
by Abbott Laboratories 510(k)
Active
510(k) K011442
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK011442 | Date5 Jun 2001 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1071-2015
One lot of alarm assemblies used in Plum A+ and Plum A+3 infusion pumps may fail to sound at all volume levels.
-
- Terminated Z-0284-2014
Plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and indicates the presence of a full or partial distal occlusion. A broken distal pin can only be detected via a visual inspection of the distal pressure pin. A broken
-
- Terminated Z-0624-2013
The door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.
-
- Terminated Z-0481-2013
The volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The manual indicates a clockwise rotation of the knob decreases the volume; however, on some of the devices the alarm volume increases when the knob is turned clockwise.
-
- Terminated Z-0204-2014
Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186, E187/N187 or E346. Additionally, if the pressure sensor calibration has drifted, users may experience false distal occlusion alarms or distal occlusions that may not
-
- Terminated Z-3284-2011
Hospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If the regulator closer does not close the flow regulator actuator when the pump door is opened, and if the clinician has not engaged the clamp prior to opening the cassette door, unrestricted flow may occur.
-
- Terminated Z-0009-2012
Hospira has received reports of continuous recycling and/or rebooting of Plum A+ infusion pumps when the backlight intensity and/or display contrast settings for the LED display has been adjusted from the original default setting. This condition could cause a delay in therapy if the clinician is unable to start the device due to the continuous recycling and/or rebooting.
-
- Terminated Z-1644-2011
Hospira has received customer reports of the Plum A+ infusion pumps with no audible alarm conditions, which have been associated with failure of the piezoelectric assembly due to component quality issues.
-
- Terminated Z-0025-2010
Fire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.
Specification
- 510(k) Number
- FDA Product Code
- FRN
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5725
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent