  "format": "devicerts.device.v1",
  "cite_as": "ABL800 Flex, Radiometer Medical ApS. deviCERTS medical device catalogue. https://devicerts.com/devices/abl800-flex-chl (data as of 2026-11-17).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/abl800-flex-chl",
    "json_url": "https://devicerts.com/devices/abl800-flex-chl.json",
    "markdown_url": "https://devicerts.com/devices/abl800-flex-chl.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-09",
    "status_as_of": "2026-11-17",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/abl800-flex-chl",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "ABL800 Flex",
    "registry_name": "ABL800 FLEX",
    "url": "https://devicerts.com/devices/abl800-flex-chl",
    "manufacturer": {
      "name": "Radiometer Medical ApS",
      "url": "https://devicerts.com/manufacturers/radiometer-medical-aps",
      "country": null,
      "country_basis": null,
      "fda_registration_country": "Denmark"
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K041874",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "CHL"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K041874",
      "number_label": null,
      "device_udi_di": null,
      "date": "2004-10-08",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0190-2027",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-08-18",
      "reason": "In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4) limitations and warnings.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-2426-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-05-31",
      "reason": "Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles.  This may result in a probability of reporting biased out-of-specification pH results on blood samples.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-1321-2022",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2022-05-18",
      "reason": "There is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-2494-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-05-26",
      "reason": "Potential risk of patient mix-up on analyzers due to software issues.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-2493-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-05-26",
      "reason": "Potential risk of patient mix-up on analyzers due to software issues.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-1684-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-03-10",
      "reason": "Analyzer's barcode reader misinterprets the contents of barcode  label used for entering patient ID or accession number. The issue is related to barcode types not using a check digit. This could result in patient mixup or loss of sample resulting in delayed medical treatment",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-1836-2020",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2020-03-06",
      "reason": "Through internal investigation is has been determined that an incorrect component was used within the power supply. The incorrect component may cause a short circuit within the power supply resulting in a delay in treatment. The delay could result in a range of risk from  cyanosis and low blood pressure  to cardiac arrhythmias and severe impairment of cardiac function.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-0682-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-07-03",
      "reason": "The analyzer software may cause a mis-match of patient demographics and test results when a certain sequence of events occur.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-0231-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-10-24",
      "reason": "Due to misinterpretation of the barcode by the scanner, when the registration receipt barcode is scanned by the analyzer , a result from a different patient will be printed or displayed on the analyzer.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-1463-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-01-26",
      "reason": "Reports that the sample type \"cord blood\" has been changed into the two sample types \"cord blood arterial\" and \"cord blood venous.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-1075-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-07-05",
      "reason": "RADIOMETER has become aware that some D788 pCO2 membranes can cause biased measurement results on patient results as well as QC results.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-2162-2012",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-07-05",
      "reason": "A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. \nIf the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. \nIf the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-2163-2012",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-07-05",
      "reason": "A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. \nIf the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. \nIf the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-1554-2009",
      "classification": "",
      "status": "terminated",
      "initiated": "2009-04-16",
      "reason": "When removing the waste bottle from the analyzer waste material may leak both from the analyzer and from the waste bottle.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-1654-2008",
      "classification": "",
      "status": "terminated",
      "initiated": "2008-01-18",
      "reason": "Reported calibration errors are not displayed on the analyzers parameter bar screen or on the patient results when the calibration number is between 32768 - 65535 and 98303 - 131070.   This may cause the operator to perform measurements and obtain questionable results.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-0038-2008",
      "classification": "",
      "status": "terminated",
      "initiated": "2007-06-22",
      "reason": "Tube may break.  The firm discovered that the yellowish/white tubes used in their blood gas analyzers (ABL800 Flex Series) were manufactured incorrectly.  These tubes have the potential to break off at the black plastic connectors.  The breakage may occur during the installation of the tubes or after a short time in use.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-1095-2007",
      "classification": "",
      "status": "terminated",
      "initiated": "2006-11-13",
      "reason": "Software anomaly. Protein removing program incorporated in the operating software for Versions: V3.833 and V5.24, of the ABL700 and ABL800 Blood Gas Analyzers, may allow for protein build up on the electrode membranes of blood gas analyzers.  This condition can influence analyzer results. When the protein removal program is run too frequently, more than once a week, or when running the protein rem",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-1325-06",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-10-31",
      "reason": "ABL800 Series Blood Gas Analyzer became inoperable when the 'restore default setup' button was touched.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-1387-06",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-09-29",
      "reason": "Under certain circumstances, the ABL 700 &800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and Sodium (Na).",
      "applies": "shared submission",
      "via": "K041874"
    },
    {
      "number": "Z-1324-06",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-09-19",
      "reason": "Reference electrodes for the firm's ABL 700/800 Blood Gas Analyzers are oversized and will not fit into the instrument.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-1319-06",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-06-17",
      "reason": "The recalled membranes cause the status value of the pH and electrolyte electrodes to be offset causing readings outside of the acceptable range.",
      "applies": "this device",
      "via": "K041874"
    },
    {
      "number": "Z-1318-06",
      "classification": "",
      "status": "terminated",
      "initiated": "2005-06-17",
      "reason": "Software defect. pO2 and pCO2 sample test results run on the firm's ABL700/800 Series Blood Gas Analyzers are not being properly flagged during the sample calibration phase.",
      "applies": "this device",
      "via": "K041874"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K041874",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "CHL",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 862.1120",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Clinical Chemistry",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}