  "format": "devicerts.device.v1",
  "cite_as": "ABL90 Flex, Radiometer Medical ApS. deviCERTS medical device catalogue. https://devicerts.com/devices/abl90-flex-chl (data as of 2026-11-17).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/abl90-flex-chl",
    "json_url": "https://devicerts.com/devices/abl90-flex-chl.json",
    "markdown_url": "https://devicerts.com/devices/abl90-flex-chl.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-09",
    "status_as_of": "2026-11-17",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/abl90-flex-chl",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "ABL90 Flex",
    "registry_name": "ABL90 FLEX",
    "url": "https://devicerts.com/devices/abl90-flex-chl",
    "manufacturer": {
      "name": "Radiometer Medical ApS",
      "url": "https://devicerts.com/manufacturers/radiometer-medical-aps",
      "country": null,
      "country_basis": null,
      "fda_registration_country": "Denmark"
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K120197",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "CHL"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K120197",
      "number_label": null,
      "device_udi_di": null,
      "date": "2012-03-19",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0010-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-25",
      "reason": "The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place:\n The analyzer is in ready mode\n Operator lifts the inlet and then closes it again, thereby initiating a rinse\n Immediately after lifts the inlet and quickly closes it again",
      "applies": "shared submission",
      "via": "K120197"
    },
    {
      "number": "Z-1046-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-12-08",
      "reason": "The ABL90 analyzer does not always use the most recent calibration data to calculate patient results. This can in some cases lead to a biased patient result.",
      "applies": "shared submission",
      "via": "K120197"
    },
    {
      "number": "Z-2164-2012",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-07-05",
      "reason": "A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. \nIf the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. \nIf the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t",
      "applies": "this device",
      "via": "K120197"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K120197",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "CHL",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 862.1120",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Clinical Chemistry",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}