# ABL90 Flex Plus

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/abl90-flex-plus-chl>
- Manufacturer: [Radiometer Medical ApS](https://devicerts.com/manufacturers/radiometer-medical-aps) (FDA registration: Denmark)
- Name in the registry: ABL90 FLEX PLUS
- Device class: Class II (US FDA)
- FDA product code: CHL

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K160153 · decision 2016-11-04 · decision SESE · status: Active

## Recalls

- Z-1659-2025 · Class II · ongoing · initiated 2025-03-21: ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.
- Z-1858-2021 · Class II · ongoing · initiated 2021-04-14 · matched through shared submission K160153, may concern other models: Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed.
- Z-1857-2021 · Class II · ongoing · initiated 2021-04-14: Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed.
- Z-1761-2020 · Class II · terminated · initiated 2020-02-03: The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.
- Z-2138-2020 · Class II · terminated · initiated 2019-11-21: The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.

## Identifiers

- 510(k) Number: K160153 [FDA 510(k)]
- FDA product code: CHL [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 862.1120 [FDA 510(k)]
- Advisory committee: Clinical Chemistry [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: ABL90 Flex Plus, Radiometer Medical ApS. deviCERTS medical device catalogue. https://devicerts.com/devices/abl90-flex-plus-chl (data as of 2026-11-17).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-11
- Statuses computed as of: 2026-11-17
- JSON: <https://devicerts.com/devices/abl90-flex-plus-chl.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/abl90-flex-plus-chl
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.