AC3 Series IABP System
510(k) K162820
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK162820 | Date31 Mar 2017 | StatusActive |
Recalls
-
- Terminated Z-2416-2020
- Terminated Z-2417-2020
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DSP
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.3535
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent