Accelerator Aps
by Abbott Laboratories 510(k)
Active
510(k) K093318
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK093318 | Date19 Feb 2010 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2251-2020
In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample dilution, which could lead to incorrect results.
-
- Terminated Z-2755-2019
The firm identified customer sites where modules which the safety pacemaker label is missing. Information regarding the label included in the Operations Manual has been updated to specify: 1) the risk of malfunction applies to anyone with a pacemaker who may work on these modules; 2) the minimum safety distance has increased to 200 mm (7.87") from transport mechanisms. Any person with a pacemaker must not get closer than this distance from modules axes.
Specification
- 510(k) Number
- FDA Product Code
- CEM
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1600
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent