Access EPO (Calibrators)
by Beckman Coulter Inc. United States EUDAMEDGUDID
Ref A16365
Product code JITEMDN W0102152206
Regulatory status
Recall ongoing on a related submission. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/IVDR EUDAMED | Number150995901AccEPOCal6F | Date2 Dec 2022 | StatusActive |
| United States | GUDID listing GUDID | Number15099590201845 | Date9 Sep 2016 | StatusIn distribution |
Coverage
- Active
- Listed
- Not covered
- European Union
- United States
Recalls
-
- Ongoing Z-1249-2025
Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased results, which may lead to improper diagnosis or repeat testing.
Specification
- Generic Name
- Erythropoietin (EPO) IVD, kit, chemiluminescent immunoassay
- Model / Ref / Catalog
- A16365
- Primary DI / UDI-DI
- 15099590201845, 15099590817404
- Basic UDI-DI
- 150995901AccEPOCal6F, 150995909AccEPOCalDP
- 510(k) Number
- FDA Product Code
- JIT
- EMDN Code
- W0102152206 Anaemia related/vitamin standards and calibrators
- GMDN Term
- Erythropoietin (EPO) IVD, kit, chemiluminescent immunoassay
- Device Class
- Class C EUDAMEDClass II GUDID
- Regulation Number
- 862.1150
- Market of Registration
- Ireland
- Commercial Distribution
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device Class | Class C | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000010288
- Authorised representative
- Beckman Coulter Ireland Inc. IE-AR-000000886
Other devices from Beckman Coulter, Inc.
Properties & raw data
Show properties, registry sources and raw records
Other properties (14)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Microbial Substances: No
- Self Testing: No
- Near-Patient Testing: No
- Companion Diagnostics: No
- Surgical Instrument: No
- Over-the-Counter: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
Sources (3)
- EUDAMED 4efeae6c-5beb-49ab-a162-a5df357f5ec7
- FDA UDI 4fb7cd0e-f5d8-4898-aff9-70a268ce8ad4
- EUDAMED 15099590817404|150995909ACCEPOCALDP