Access Intact PTH Reagent Pack
Ref A16972
Active
CE IVDR 510(k)
Valid until
Not listed
FDA recalls
1 ongoingTrust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k) 510(k) | NumberK232791 | Date1 Mar 2024 | StatusActive |
| European Union | CE/IVDR EUDAMED | Number150995901AccInPTHQC 15099590201937 | Date2 May 2022 | StatusActive |
| United States | GUDID listing GUDID | Number15099590201937 | Date9 Sep 2016 | StatusListed |
Certificates
EUDAMED lists no certificate for this device's Basic UDI-DI.
Recalls
-
- Ongoing Z-0673-2025
The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the intact PTH (parathyroid hormone) test may produce falsely decreased intact parathyroid hormone results. This could lead to a missed diagnosis, a delay in treatment, or inappropriate clinical management, depending upon the patient s clinical condition.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- United States
- European Union
Specification
- Trade Name
- Access Intact PTH Reagent Pack EUDAMEDAccess Intact PTH GUDID
- Generic Name
- Parathyroid hormone (PTH) IVD, kit, chemiluminescent immunoassay
- Model / Ref / Catalog
- A16972
- Primary DI / UDI-DI
- 15099590201937
- Basic UDI-DI
- 150995901AccInPTHQC
- FDA Product Code
- CEW
- EMDN Code
- W0102060312 Parathyroid hormone intact
- GMDN Term
- Parathyroid hormone (PTH) IVD, kit, chemiluminescent immunoassay
- Device Class
- Class B EUDAMEDClass II GUDID
- Regulation Number
- 862.1545
- EU market of registration
- Ireland
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device Class | Class B | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000010288
- Authorised representative
- Beckman Coulter Ireland Inc. IE-AR-000000886
Other devices from Beckman Coulter, Inc.
Properties
Other properties (16)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Human Tissues: No
- Microbial Substances: No
- Self Testing: No
- Near-Patient Testing: No
- Companion Diagnostics: No
- Surgical Instrument: No
- New Device: No
- Over-the-Counter: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No