Access Total T3 Reagents on the Access Immunoassay Analyzer
by Beckman Instruments, Inc. 510(k)
Active
510(k) K973966
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK973966 | Date31 Oct 1997 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2239-2020
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.
-
- Terminated Z-2635-2017
The Access immunoassays are susceptible to biotin interference.
-
- Terminated Z-2227-2012
A recall was initiated because Beckman Coulter has identified a negative bias in test results when comparing elevated Total T3 patient samples diluted with Sample Diluent A to samples diluted with Access Total T3 Calibrator S0.
Specification
- 510(k) Number
- FDA Product Code
- CDP
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1710
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 26 Sep 2026
The Access Total T3 Reagents on the Access Immunoassay Analyzer is a paramagnetic particle chemiluminescent immunoassay for the quantitative determination of triiodothyronine levels in human serum and plasma. It is classified as a Radioimmunoassay, Total Triiodothyronine device under product code CDP and device class 2.
The device is supplied for use with the Access Immunoassay Systems and was cleared via FDA 510(k) submission K973966 on October 31, 1997, as substantially equivalent to a predicate device. It is intended for in vitro diagnostic use in clinical chemistry settings.
Other devices from Beckman Instruments, Inc.
Questions
AI-generated answers, 26 Sep 2026
What is the Access Total T3 Reagents used for?
It is used for the quantitative determination of triiodothyronine (T3) levels in human serum and plasma as an in vitro diagnostic test in clinical chemistry settings.
How is the Access Total T3 Reagents supplied and what system is it designed for?
The reagents are supplied for use with the Access Immunoassay Systems, specifically the Access Immunoassay Analyzer, and are intended for in vitro diagnostic use in clinical laboratories.
Is the Access Total T3 Reagents a single-use or reusable device?
As an in vitro diagnostic reagent kit, it is intended for single-use in each test run; the reagents are consumed during the assay process and not reused.