Access Ultrasensitive Insulin Assay
by Bio-Rad Laboratories, Inc. 510(k)
Active
510(k) K953084
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK953084 | Date25 Sep 1995 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-0981-2024
There is an issue with the Access Ultrasensitive Insulin assay protocol file (APF) for use with the DxI 9000 Access Immunoassay Analyzer (APF 182 test definition version 6.7 and below). The Access Ultrasensitive Insulin APF applies an incorrect conversion factor when converting from system default units of IU/mL to the International System of Units (SI units) of pmol/L.
-
- Terminated Z-1160-2008
False negative results when used to test serum samples (as compared to plasma samples).
Specification
- 510(k) Number
- FDA Product Code
- CFP
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1405
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent