# Accu-Check Combo System

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/accu-check-combo-system-lzg>
- Manufacturer: [Roche Diagnostics](https://devicerts.com/manufacturers/roche-diagnostics-corp) (United States, FDA 510(k) applicant)
- Name in the registry: ACCU-CHECK COMBO SYSTEM
- Device class: Class II (US FDA)
- FDA product code: LZG

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K111353 · decision 2012-07-17 · decision SESE · status: Active

## Recalls

- Z-1733-2015 · Class II · terminated · initiated 2015-04-22: Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and El0 error messages. If the user does not act upon the E6 and E10 error messages appropriately, insulin delivery will be stopped and if unnoticed may lead to severe hyperglycemia u
- Z-0146-2015 · Class II · terminated · initiated 2014-09-25: The insulin pump may lose time and date settings during a power interruption (such as a battery change) due to a faulty capacitor. If the capacitor fails, the time and date will return to default setting. Though the pump prompts customers to confirm the time and date, if the user does not recognize the time and date have returned to default, a shift of their basal rate time block could occur.
- Z-0385-2014 · Class II · terminated · initiated 2013-10-25 · matched through shared submission K111353, may concern other models: There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pump option is chosen on the ACCU-CHEK Aviva Combo meter, there is a possibility of receiving an incorrect bolus advice recommendation that may cause a temporary under delivery of insulin.

## Identifiers

- 510(k) Number: K111353 [FDA 510(k)]
- FDA product code: LZG [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 880.5725 [FDA 510(k)]
- Advisory committee: General Hospital [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Accu-Check Combo System, Roche Diagnostics. deviCERTS medical device catalogue. https://devicerts.com/devices/accu-check-combo-system-lzg (data as of 2026-11-24).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-24
- JSON: <https://devicerts.com/devices/accu-check-combo-system-lzg.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/accu-check-combo-system-lzg
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.