Accu-Chek Advantage Test Strips and Accu-Chek Advantage H Test Strips
510(k) K982089
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK982089 | Date3 Sep 1998 | StatusActive |
Recalls
-
- Terminated Z-1378-06
The drying agent beads, located in the cap, may become loose and fall into the vial causing the test strips to potentially give incorrect blood glucose test results
-
- Terminated Z-0569-06
- Terminated Z-0567-06
- Terminated Z-0570-06
- Terminated Z-0568-06
There may be a thin piece of plastic across the mouth of the vial, preventing the cap from sitting snugly onto the vial and, if present, this situation will cause erroneously low or high blood glucose readings.
-
- Terminated Z-0573-06
There may be a thin piece of plastic across the mouth of the vial, preventing the cap from sitting snugly onto the vial and, if present, this situation will cause erroneously low or high blood glucose readings..
-
- Terminated Z-0716-03
Crack in the bottom of the test strip vial will cause erroneously low or high blood glucose readings.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LFR
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1345
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent