Accudata Glucose Test Station
510(k) K924475
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK924475 | Date13 Oct 1992 | StatusActive |
Recalls
-
- Terminated Z-1368-06
- Terminated Z-1366-06
- Terminated Z-1367-06
The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.
-
- Terminated Z-0883-05
- Terminated Z-0881-05
- Terminated Z-0882-05
If foil wrapped packages or other metallic objects are lodged behind the supply drawer, an electrical short may result and high heat generated; this could result in plastic melting and battery leakage for customers using 'C' cell batteries to power their base and could render the system inoperable (system won't turn on) for customers using the A/C adaptor without batteries.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CGA
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1345
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent