Achieva 1.5T & Inera 1.5T Family
by Philips Medical Systems Nederland B.V.
510(k) K052013
Active
Valid until
No expiry
FDA recalls
10 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK052013 | Date2 Aug 2005 | StatusActive |
Recalls
-
- Ongoing Z-3232-2026
- Ongoing Z-3240-2026
- Ongoing Z-3233-2026
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
-
- Ongoing Z-2165-2024
- Ongoing Z-2163-2024
- Ongoing Z-2164-2024
Potential for coils to heat up and harm patients (burn).
-
- Ongoing Z-0393-2024
- Ongoing Z-0392-2024
- Ongoing Z-0389-2024
- Ongoing Z-0390-2024
Potential component failure in the Gradient Coil could product smoke and/or fire.
-
- Terminated Z-0888-2019
- Terminated Z-0891-2019
- Terminated Z-0889-2019
- Terminated Z-0887-2019
- Terminated Z-0909-2019
- Terminated Z-0894-2019
- Terminated Z-0895-2019
- Terminated Z-0892-2019
- Terminated Z-0915-2019
- Terminated Z-0912-2019
- Terminated Z-0908-2019
- Terminated Z-0899-2019
- Terminated Z-0890-2019
- Terminated Z-0896-2019
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
-
- Terminated Z-1433-2018
- Terminated Z-1435-2018
Potential risk for helium gas inside the MR examination room during a magnet quench
-
- Terminated Z-2461-2010
The cables of the coil may become too hot and cause burns when placed too close to the patient. While using the 1.5T Sense XL Torso coil, unwanted radiofrequencies (RF) interaction with the patient can occur. Especially in combination with high SAR levels and when the RF cables of the coil are too close to the patient, this RF interaction has resulted in RF burns to patients.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LNH
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1000
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent
Other devices from Philips Medical Systems Nederland B.V.
See all devices from Philips Medical Systems Nederland B.V. →