Acs FRT4 Immunoassay

by Ciba Corning Diagnostics Corp. 510(k)

Active United States

510(k) K961510

Regulatory status

as of 29 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK961510 Date17 Jun 1996ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

4 recalls · FDA

  • 3 recalls, same reason

    • Terminated Z-1254-2017 FDA Class II initiated 10 Jan 2017
    • Terminated Z-1256-2017 FDA Class II initiated 10 Jan 2017
    • Terminated Z-1255-2017 FDA Class II initiated 10 Jan 2017

    Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur FT4 when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT Systems. In addition, Siemens Healthcare Diagnostics confirmed the potential for calibration failures due to above limit calibrator RLU %CVs when using Calibrator A kit lots ending in 90 with the FT4 assay. The performance of the FT4 assay when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur CP System is not affected. Customers may continue to use Calibrator A kit lots ending in 90 for the ADVIA Centaur Systems FT3, T3, T4 and TUp assays.

    • Terminated Z-1970-2013 FDA Class III initiated 28 Jun 2013 on shared 510(k) K961510

    Calibrator A lot CA77 is under-recovering to the assigned target for the ADVIA Centaur Total T4 Assay.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Specification

510(k)

510(k) Number
FDA Product Code
JJX
Device Class
Class I
Regulation
21 CFR Part 862.1660