Acs FRT4 Immunoassay
by Ciba Corning Diagnostics Corp. 510(k)
Active
510(k) K961510
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK961510 | Date17 Jun 1996 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1254-2017
- Terminated Z-1256-2017
- Terminated Z-1255-2017
Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur FT4 when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT Systems. In addition, Siemens Healthcare Diagnostics confirmed the potential for calibration failures due to above limit calibrator RLU %CVs when using Calibrator A kit lots ending in 90 with the FT4 assay. The performance of the FT4 assay when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur CP System is not affected. Customers may continue to use Calibrator A kit lots ending in 90 for the ADVIA Centaur Systems FT3, T3, T4 and TUp assays.
-
- Terminated Z-1970-2013
Calibrator A lot CA77 is under-recovering to the assigned target for the ADVIA Centaur Total T4 Assay.
Specification
- 510(k) Number
- FDA Product Code
- JJX
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1660
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent