Active Inhibin-a Elisa, Model Dsl-10-28100

by Diagnostic Systems Laboratories, Inc. 510(k)

Active United States

510(k) K002128

Regulatory status

as of 29 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK002128 Date5 Dec 2000ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

4 recalls · FDA

  • 4 recalls, same reason

    • Terminated Z-2774-2011 FDA Unclassified (pre-2012) initiated 31 Mar 2011
    • Terminated Z-2772-2011 FDA Unclassified (pre-2012) initiated 31 Mar 2011
    • Terminated Z-2775-2011 FDA Unclassified (pre-2012) initiated 31 Mar 2011
    • Terminated Z-2773-2011 FDA Unclassified (pre-2012) initiated 31 Mar 2011

    The recall was initiated because patient results are falsely elevated by approximately 30% because the reference standard of the Manual Inhibin A ELISA assay has shifted from the assays original calibration. Erroneously elevated results may lead to unnecessary ultrasound testing to confirm gestational age, patient psychological stress until diagnostic testing is completed, and/or medical decis

Specification

510(k)

510(k) Number
FDA Product Code
NDR
Device Class
Class I
Regulation
21 CFR Part 862.1260
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent