Adapter, AC, CADD-SOLIS
Ref 21-0270-25
Active
Valid until
Not listed
FDA recalls
5 ongoingDocuments
None linked in EUDAMED
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number0610586M0404D8 15019517126884 | Date8 Oct 2025 | StatusActive |
| United States | GUDID listing GUDID | Number15019517126884 | Date3 Jul 2018 | StatusListed |
Certificates
EUDAMED lists no certificate for this device's Basic UDI-DI.
Recalls
-
- Ongoing Z-1682-2025
There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.
-
- Ongoing Z-1668-2025
Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
-
- Ongoing Z-0222-2025
Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable.
-
- Ongoing Z-0466-2025
The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or more of the AC mains prongs may separate from the input plug.
-
- Ongoing Z-2421-2024
The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.
-
- Terminated Z-0107-2022
Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line event may not be recognized by the pump and may not alarm to notify the clinician.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade Name
- Adapter, AC, CADD-SOLIS EUDAMEDCadd GUDID
- Generic Name
- Basic power supply, reusable
- Device Name
- CADD-Solis AC Adapter
- Model / Reference
- 21-0270-25
- Primary DI / UDI-DI
- 15019517126884
- Basic UDI-DI
- 0610586M0404D8
- 510(k) Number
- FDA Product Code
- FRN
- EMDN Code
- Z12030380 Infusion instruments - hardware accessories
- GMDN Term
- Basic power supply, reusable
- Device Class
- Class I EUDAMEDClass II GUDID
- Regulation Number
- 880.5725
- EU market of registration
- Netherlands
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device Class | Class I | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000002566
- Authorised representative
- ICU Medical BV NL-AR-000002748
Other devices from ICU Medical, Inc.
Properties
Other properties (22)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Prescription Use: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No