Skip to content

Add-A-Vial II Binary Connector

by Kendall Mcgaw Laboratories, Inc.

510(k) K900865

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK900865 Date11 Jul 1990ClearedStatusActive

Recalls

2 recalls · FDA

  • 2 recalls, same reason

    • Terminated Z-0424-2014 FDA Class II initiated 4 Apr 2013
    • Terminated Z-0425-2014 FDA Class II initiated 4 Apr 2013

    Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LHI
Device Class
Class II
Regulation
21 CFR Part 880.5440
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent

Other devices from Kendall Mcgaw Laboratories, Inc.

See all devices from Kendall Mcgaw Laboratories, Inc. →