Advance Total Knee System
by Wrightmedicaltechnologyinc 510(k)
Active
510(k) K061223
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK061223 | Date18 Aug 2006 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2696-2016
- Terminated Z-2695-2016
- Terminated Z-2697-2016
- Terminated Z-2701-2016
- Terminated Z-2703-2016
- Terminated Z-2698-2016
- Terminated Z-2702-2016
- Terminated Z-2694-2016
- Terminated Z-2699-2016
- Terminated Z-2693-2016
- Terminated Z-2700-2016
ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.
-
- Terminated Z-0492-2010
- Terminated Z-0493-2010
Titanium used in the recalled devices was discovered to have internal microstructure conditions of alloy segregation, primary alpha and/or a transformed beta matrix in some of the heat lots. The absence of alloying elements results in an area with less strength. For those instances where the product has been implanted, there is the potential increased risk of failure for these patients.
Specification
- 510(k) Number
- FDA Product Code
- MBH
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3565
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent