  "format": "devicerts.device.v1",
  "cite_as": "Advanced Perfusion System 1, Terumo Cardiovascular Systems Corporation. deviCERTS medical device catalogue. https://devicerts.com/devices/advanced-perfusion-system-1-dtq (data as of 2026-11-26).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/advanced-perfusion-system-1-dtq",
    "json_url": "https://devicerts.com/devices/advanced-perfusion-system-1-dtq.json",
    "markdown_url": "https://devicerts.com/devices/advanced-perfusion-system-1-dtq.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-11",
    "status_as_of": "2026-11-26",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/advanced-perfusion-system-1-dtq",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Advanced Perfusion System 1",
    "registry_name": "Advanced Perfusion System 1",
    "url": "https://devicerts.com/devices/advanced-perfusion-system-1-dtq",
    "manufacturer": {
      "name": "Terumo Cardiovascular Systems Corporation",
      "url": "https://devicerts.com/manufacturers/terumo-cardiovascular-systems-corporation",
      "country": null,
      "country_basis": null,
      "fda_registration_country": "United States"
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K172220",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "DTQ"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K172220",
      "number_label": null,
      "device_udi_di": null,
      "date": "2017-08-17",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-2350-2021",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2021-06-08",
      "reason": "intermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs",
      "applies": "this device",
      "via": "K172220"
    },
    {
      "number": "Z-2959-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-08-25",
      "reason": "Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector.  This intermittent connection may result in false air bubble alarms.",
      "applies": "this device",
      "via": "K172220"
    },
    {
      "number": "Z-2960-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-08-25",
      "reason": "Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector.  This intermittent connection may result in false air bubble alarms.",
      "applies": "this device",
      "via": "K172220"
    },
    {
      "number": "Z-2961-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-08-25",
      "reason": "Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector.  This intermittent connection may result in false air bubble alarms.",
      "applies": "this device",
      "via": "K172220"
    },
    {
      "number": "Z-1108-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-01-10",
      "reason": "These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed.  The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.",
      "applies": "this device",
      "via": "K172220"
    },
    {
      "number": "Z-1109-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-01-10",
      "reason": "These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed.  The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.",
      "applies": "this device",
      "via": "K172220"
    },
    {
      "number": "Z-1110-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-01-10",
      "reason": "These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed.  The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.",
      "applies": "this device",
      "via": "K172220"
    },
    {
      "number": "Z-1111-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-01-10",
      "reason": "These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed.  The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.",
      "applies": "this device",
      "via": "K172220"
    },
    {
      "number": "Z-1337-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-04-18",
      "reason": "The gas flow rate output of the EPGS may be inaccurate due to a defect in the internal flowmeter.",
      "applies": "this device",
      "via": "K172220"
    },
    {
      "number": "Z-1769-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-10-17",
      "reason": "An intermittent failure which causes a blender initialization fault upon start-up or during an EPGS calibration attempt leads to the Central Control Monitor control sliders being disabled, such that the gas system can only be controlled by the local control knobs.",
      "applies": "this device",
      "via": "K172220"
    },
    {
      "number": "Z-0176-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-09-13",
      "reason": "The service manual was not updated at the time of the release of software version 1.30.",
      "applies": "this device",
      "via": "K172220"
    },
    {
      "number": "Z-3242-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-08-28",
      "reason": "Screws on a Terumo Advanced Perfusion\nSystem 1 Power Manager Board (the base unit circuit board) were not torqued to specification.  It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system.",
      "applies": "this device",
      "via": "K172220"
    },
    {
      "number": "Z-3243-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-08-28",
      "reason": "Screws on a Terumo Advanced Perfusion\nSystem 1 Power Manager Board (the base unit circuit board) were not torqued to specification.  It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and  may loosen over time, resulting in the potential for an electrical short in the system",
      "applies": "this device",
      "via": "K172220"
    },
    {
      "number": "Z-1478-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-01-19",
      "reason": "Inaccurrate flow readings.  Depending on the degree of inaccuracy, this issue may not be easy for the user to detect during setup or use (for example, following a Flow Probe relocation or manipulation).",
      "applies": "this device",
      "via": "K172220"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K172220",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "DTQ",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 870.4220",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Cardiovascular",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}