Skip to content

Advantx Lcv+

by GE Medical Systems

510(k) K960575

Advantx Lcv+ is a Class II device made by GE Medical Systems, cleared under FDA 510(k) K960575 on 7 May 1996, active as of 22 Oct 2026. 2 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

6 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK960575 Date7 May 1996ClearedStatusActive

Official records

Recalls

6 recalls · FDA

    • Terminated Z-0908-2017 FDA Class II initiated 21 Nov 2016

    GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors - Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems.

    • Terminated Z-0763-2016 FDA Class II initiated 18 Dec 2015 on shared 510(k) K960575

    GE Healthcare has recently become aware of a reported incident in which a Mavig suspension arm with monitor assembly fell to the floor.

    • Terminated Z-0021-2014 FDA Class II initiated 16 Aug 2013 on shared 510(k) K960575

    GE Healthcare has become aware of a potential safety issue involving missing screws in the overhead Video Monitor Suspension of the Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, Innova 4100IQ Cardiovascular X-ray imaging systems.

    • Terminated Z-0616-2010 FDA Unclassified (pre-2012) initiated 22 Dec 2009 on shared 510(k) K960575

    Potential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some table rotational brakes, an unintended table rotation may occur while the table is locked. This issue could lead to a patient fall if the unintended motion occurs during patient

    • Terminated Z-2359-2008 FDA Unclassified (pre-2012) initiated 24 Jun 2008 on shared 510(k) K960575

    GE Healthcare has become aware of 2 potential conditions that could occur with the MAVIG GD Monitor Suspension that is used with Advantx-E, Innova 2000, 21000IQ, 3100/3100IQ, 4100/4100IQ and 2121 IQ/3131IQ cardiovascular systems. The two conditions include 1). Due to insufficient securing of the connecting elements, the LCD vertical monitor support may disengage from its arm and fall on the tab

    • Terminated Z-2038-2008 FDA Unclassified (pre-2012) initiated 14 Jan 2008

    Risk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

7 dated events

  1. Cleared FDA 510(k) K960575
  2. Recall initiated FDA recalls Z-2038-2008
  3. Recall initiated FDA recalls Z-2359-2008
  4. Recall initiated FDA recalls Z-0616-2010
  5. Recall initiated FDA recalls Z-0021-2014
  6. Recall initiated FDA recalls Z-0763-2016
  7. Recall initiated FDA recalls Z-0908-2017

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
IZI
Device class
Class II
Regulation
21 CFR Part 892.1600
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code IZI (System, X-Ray, Angiographic).

Other devices from GE Medical Systems, Inc.

See all devices from GE Medical Systems, Inc. →