Advia 120 Hematology System, in Vitro Diagnostic System
by Bayer Corp.
510(k) K971998
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK971998 | Date29 Aug 1997 | StatusActive |
Recalls
-
- Terminated Z-2114-2013
The ADVIA 120 optics cover or hood is located on the very top of the ADVIA 120 analyzer. When the optics cover is raised, it is supported by two gas struts (SMN 10309266) attached to the cover on the left and right sides. Over time, it is possible that these struts may lose their effectiveness and fail to support the optics cover in the open position. This could cause the cover to fall during maintenance procedures.
-
- Terminated Z-1376-06
The ADVIA 2120 has reported ; highly intermittent low results on all primary results: White Blood Cells (WBC), Red Blood Cells (RBC), Hemoglobin (HGB) and Platelets (PLT) and an abnormal baso cytogram.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- GKL
- Device Class
- Class II
- Regulation
- 21 CFR Part 864.5200
- Advisory Committee
- Hematology
- FDA Decision
- SESE Substantially Equivalent