Advia Cantaur and ACS:180 Insulin Calibrator
510(k) K021535
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK021535 | Date27 Jun 2002 | StatusActive |
Recalls
-
- Terminated Z-2085-2017
- Terminated Z-2086-2017
Current lots of ADVIA Centaur/XP/XPT/CP Insulin (IRI) assay recover approximately 40% higher than the World Health Organization (WHO) 1st IRP 66/304 standardization based on slope values. Other performance characteristics are being met per the assay Instructions for Use. Overall risk to health is negligible and Siemens is not recommending a review of previously generated results.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JIT
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1150
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent