Advia Chemistry TRIGLYCERIDES_2 (TRIG_2) Reagent
by Siemens Healthcare Diagnostics, Inc.
510(k) K133067
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK133067 | Date22 Nov 2013 | StatusActive |
Recalls
-
- Ongoing Z-1270-2023
Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results
-
- Terminated Z-1657-2016
Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific
-
- Terminated Z-1670-2016
Siemens Healthcare Diagnostics has confirmed the ADVIA Chemistry Triglyceride concentrated reagent (TRIG_c) does not meet Instructions For Use (IFU) linearity claims at the upper limit of the assay range as it approaches end of shelf life for lots # 348297 and 359932.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CDT
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1705
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent