Recalls
-
- Ongoing Z-1237-2026
A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.
-
- Ongoing Z-0113-2024
One lot of Tournier Perform Reversed Lateralized Baseplates were found to have been manufactured with 6mm spacers instead of 3mm spacers. This nonconformance prevents the central screw from being able to thread into the baseplate.
-
- Ongoing Z-1216-2022
The device is unable to meet sterilization requirements when sterilized per instructions in an Aesculap container.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- AEQUALIS™ PerFORM Reversed
- Generic Name
- Orthopaedic bone screw (non-sliding)
- Model / Reference
- DWJ625
- Primary DI / UDI-DI
- 00846832061082
- FDA Product Code
- PHX, KWS
- GMDN Term
- Orthopaedic bone screw (non-sliding)
- Device Class
- Class II
- Regulation Number
- 888.3660
- GUDID distribution status
- Not in Commercial Distribution
- Size
- Outer Diameter — 9.5 MillimeterLength — 25 Millimeter
Other devices from Tornier, Inc.
Properties
Other properties (8)
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Manufacturing Date: No