Aeroset System
by Abbott Laboratories 510(k)
Active
510(k) K980367
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK980367 | Date1 Apr 1998 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1671-2020
Mixer blade may separate from mixer due to the screw and nut failure and detachment. As a result of the mixer blade missing, the sample and reagent are inadequately mixed.
-
- Terminated Z-1296-2019
There is a potential to generate falsely elevated serum or plasma chloride results when using the product.
-
- Terminated Z-1104-2019
- Terminated Z-1103-2019
- Terminated Z-1102-2019
There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
-
- Terminated Z-0921-2018
- Terminated Z-0922-2018
- Terminated Z-0923-2018
There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000, c8000, and c16000 instrument.
-
- Completed Z-0165-2018
- Completed Z-0166-2018
The Peristaltic Head tubing placed on the ARCHITECT c4000, c8000, and c16000 has the possibility to leak due to a manufacturing issue associated with the tubing connector.
-
- Terminated Z-0163-2016
Incorrect tubing.
-
- Terminated Z-2202-2015
The ARCHITECT c8000 instrument contains tubing that does not meet specifications and has the potential to leak.
-
- Terminated Z-1725-2015
The product is leaking from the cap and crystallization was noted as present around the cap.
-
- Terminated Z-0011-2018
There is a potential to generate falsely-depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000 instrument.
-
- Terminated Z-1932-2015
ICT Serum Calibrator may generate lower than expected Potassium Quality Control (QC) and patient results on the Abbott Architect cSystem.
-
- Terminated Z-2231-2008
Software defect can allow tests ordered for one sample to be aspirated from a different sample, reporting erroneous results for the affected sample.
-
- Terminated Z-1166-2008
Quality Control (Out of Range)- ICT Calibrator lot number: 0505017, may contain an interferent. Not all vials within a kit are affected. The investigation determined that the interferent affects Potassium (K+) only. Use of the affected calibrator vials will generate an acceptable, but low calibration slope. When Quality Control (QC) is run to verify the calibration, K+ QC results may be bel
-
- Terminated Z-0146-2008
Incorrect Sample/Test Identification: Due to a software timing defect with certain versions of ARCHITECT cSystem Software, test results may be incorrectly associated to a Sample Identification during specific sample-processing conditions of the Robotic Sample Handler.
-
- Terminated Z-0952-2007
One or more of four software issues : 1) calculation error for Architect cSystem assays using the Spline math model. 2) The system configuration option, Run controls onboard reagents by Kit, does not function correctly for assays requiring a standard sample dilution. 3) Infrequent data may be missing from automatically or manually printed reports generated on the Architect system. 4) Clinical Chem
-
- Terminated Z-0398-06
System software assigns a calibrator default volume of 2.0uL when field is left empty by operator at time assay parameters set. Patient results could be affected if the volume required is not 2.0 uL.
-
- Terminated Z-0010-06
Software deficiency causing potential shift in absorbance readings for test results.
-
- Terminated Z-0008-06
Device has incorrect notation associated with therapeutic drug monitoring assays.
-
- Terminated Z-0016-05
Changes to the printer default settings made by users/operators cause printed data to be truncated and misidentified.
-
- Terminated Z-1341-04
Process change in the manufacturing of the AC/DC driver board used on the analyzer may result in one or more of the solid state relays to fail during use.
-
- Terminated Z-0846-04
Change of level 3 error log messages to level 1 to prevent suspect assay results.
-
- Terminated Z-0818-04
When using a LIS the software system will send up a test result of up to 20 characters although the field is set to display only 15 characters per the RS-232 Manual.
-
- Terminated Z-0263-04
Sample Identification (SID) numbers of a bar coded sample may be truncated, or shortened, after being scanned by the bar code reader (e.g. an actual SID of 123456 is shortened to 1234)
-
- Terminated Z-0091-04
Audible alarm occurs before aspiration potentially resulting in operator injury or results assigned to wrong patient for STAT samples.
-
- Terminated Z-0788-03
Software failure to detect error in reference solution set-up can cause inaccurate results for Sodium, Potassium, and Chloride.
-
- Terminated Z-0725-03
Under certain operating conditions, smart wash feature disables with multiple samples.
-
- Terminated Z-0796-03
External waste pump has metal edge around perimeter sharp enough to cause scratch or cut.
-
- Terminated Z-0844-03
Design of high concentration waste line causes formation of a precipitate in the waste line causing overflow of waste.
-
- Terminated Z-0843-03
Extensive use of cuvettes causes inaccurate results. Preventative maintenance did not specify replacement schedule.
-
- Terminated Z-1142-03
ICT assays run using a manual dilution will not be calculated correctly by the software.
Specification
- 510(k) Number
- FDA Product Code
- CGZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1170
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent