Agilis Steerable Catheter Introducer, Model 408304
by St. Jude Medical, Daig Division, Inc.
510(k) K042623
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK042623 | Date27 Jan 2005 | StatusActive |
Recalls
-
- Terminated Z-1086-06
St Jude Medical has determined that a limited number of Agilis Steerable Introducers manufactured in late 2005 are susceptible to the hemostasis hub leaking or separating from the handle. The risks that exist are related to the loss of hemostasis through the Agilis introducer and/or the possibility of air or foreign body being introduced into the patient with resultant embolism.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DYB
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.1340
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent