Alere Cholestech Ldx Analyzer, Alere Cholestech Ldx Lipid Profile - Glu Cassette
by Alere San Diego Inc. Dba Biosite, Innovacon, Hemosen
510(k) K120615
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK120615 | Date15 May 2013 | StatusActive |
Recalls
-
- Ongoing Z-1702-2024
There is a potential that the Cholestech LDX battery (an external battery pack available as an optional accessory to the Cholestech LDX Analyzer) may swell which could cause minor burns, cuts or abrasions, or low current electrical shock to the user.
-
- Ongoing Z-1508-2022
Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).
-
- Terminated Z-2475-2019
The polarity of the output of power supply was reversed. Due to polarity of the output of power supplies being reversed, it renders the power supplies nonfunctional and the analyzer inoperable.
-
- Terminated Z-2778-2016
Incorrect expiration date on the magnetic strip that does not match the labeled expiration date.
-
- Terminated Z-0379-2014
Alere initiated a recall of Cholestech LDX Lipid Profile-GLU test cassettes, Catalog Number 10-991, Lot # 320167, becuase an incorrect lot number was printed on the outer 10 pack kit box label. The incorrect Lot # number printed was 321067.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CGA
- FDA applicant
- Alere San Diego Inc. Dba Biosite,Innovacon,Hemosen
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1345
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent