Alinity h-series System
by Abbott Laboratories 510(k)
Active
510(k) K220031
Regulatory status
Recall ongoing. 2 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK220031 | Date4 Aug 2023 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-0489-2026
Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.
-
- Ongoing Z-2681-2024
1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating falsely high Mean Cell Hemoglobin(MCH)/MCHC and falsely low hematocrit results. 2)Cell events may be incorrectly counted as basophil(BASO), resulting in increased BASO and %BASO counts
Specification
- 510(k) Number
- FDA Product Code
- GKZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 864.5220
- Advisory Committee
- Hematology
- FDA Decision
- SESE Substantially Equivalent