Alinity m 09N15-03C
Listed
Valid until
Not applicable
FDA recalls
1 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00884999051928 | Date27 Jul 2026 | StatusListed |
Recalls
-
- Ongoing Z-0233-2025
Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- Alinity m
- Generic Name
- Laboratory instrument/analyser application software IVD
- Model / Ref / Catalog
- 09N15-03C
- Primary DI / UDI-DI
- 00884999051928
- PMA Number
- FDA Product Code
- MAQ
- GMDN Term
- Laboratory instrument/analyser application software IVD
- Device Class
- Class III
- GUDID distribution status
- In Commercial Distribution
Other devices from Abbott Molecular
Properties
Other properties (12)
- Single Use: No
- Implantable: No
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No