Skip to content

Allon 2001 Version 5

by Mtre Advanced Technology, Ltd.

510(k) K024128

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK024128 Date10 Feb 2003ClearedStatusActive

Recalls

1 recall · 1 ongoing · FDA

    • Ongoing Z-1966-2025 FDA Class II initiated 13 May 2025

    An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
DWJ
FDA applicant
Mtre Advanced Technology , Ltd.
Device Class
Class II
Regulation
21 CFR Part 870.5900
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent

Other devices from Mtre Advanced Technology, Ltd.

  • Allon 2000

    Class II
  • Criticool Open recall

    Class II
  • Criticool Pro

    Class II
  • Allon 2001 Modified

    Class II
  • Allon 2001

    Class II