Allura Xper FD series and Allura Xper OR Table series
by Philips Medical Systems Nederland B.V. 2 variants
510(k) K162859
Active
Valid until
No expiry
FDA recalls
36 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK162859 | Date2 Dec 2016 | StatusActive |
Recalls
-
- Ongoing Z-2928-2026
Firmware issue potentially leading to loss of geometry movements.
-
- Ongoing Z-2503-2025
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
-
- Ongoing Z-2189-2025
Potential for temporary loss of imaging (X-ray) functionality due to software issue.
-
- Ongoing Z-1080-2025
- Ongoing Z-1084-2025
- Ongoing Z-1082-2025
- Ongoing Z-1079-2025
- Ongoing Z-1086-2025
- Ongoing Z-1083-2025
- Ongoing Z-1081-2025
- Ongoing Z-1077-2025
- Ongoing Z-1085-2025
- Ongoing Z-1078-2025
- Ongoing Z-1076-2025
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
- Ongoing Z-1146-2025
- Ongoing Z-1152-2025
- Ongoing Z-1147-2025
- Ongoing Z-1148-2025
- Ongoing Z-1143-2025
- Ongoing Z-1161-2025
- Ongoing Z-1151-2025
- Ongoing Z-1144-2025
- Ongoing Z-1149-2025
- Ongoing Z-1142-2025
- Ongoing Z-1150-2025
- Ongoing Z-1145-2025
Possibility of the patient falling from the table related to the mattress used on the patient table.
-
- Ongoing Z-1895-2024
- Ongoing Z-1896-2024
- Ongoing Z-1897-2024
When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
-
- Ongoing Z-1156-2024
Dual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
-
- Ongoing Z-1144-2024
Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
-
- Ongoing Z-1151-2024
Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
-
- Ongoing Z-0558-2024
Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose
-
- Ongoing Z-2587-2023
A damaged foot switch may cause instances where no or only intermittent x-ray radiation initiation is possible.
-
- Ongoing Z-2402-2023
A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation
-
- Ongoing Z-2283-2023
The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- OWB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent