Allura Xper FD10
by Philips Medical Systems North America Co.
510(k) K041949
Active
Valid until
No expiry
FDA recalls
16 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK041949 | Date30 Jul 2004 | StatusActive |
Recalls
-
- Ongoing Z-0007-2026
- Ongoing Z-0012-2026
- Ongoing Z-0011-2026
- Ongoing Z-0009-2026
- Ongoing Z-0006-2026
- Ongoing Z-0005-2026
- Ongoing Z-0004-2026
- Ongoing Z-0001-2026
- Ongoing Z-0008-2026
- Ongoing Z-0002-2026
- Ongoing Z-0013-2026
- Ongoing Z-0014-2026
- Ongoing Z-0003-2026
- Ongoing Z-0015-2026
- Ongoing Z-0010-2026
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
- Ongoing Z-1077-2025
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
- Terminated Z-1140-2019
- Terminated Z-1141-2019
Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.
-
- Terminated Z-3227-2018
- Terminated Z-3225-2018
- Terminated Z-3229-2018
- Terminated Z-3226-2018
- Terminated Z-3230-2018
- Terminated Z-3228-2018
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
-
- Terminated Z-1822-2017
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
-
- Terminated Z-1062-2017
- Terminated Z-1061-2017
System may lock up.
-
- Terminated Z-0923-2017
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.
-
- Terminated Z-0810-2016
The actuator assembly became detached and the monitor carriage with a flexvision monitor dropped to its lowest position.
-
- Terminated Z-0354-2016
- Terminated Z-0352-2016
- Terminated Z-0346-2016
- Terminated Z-0349-2016
- Terminated Z-0353-2016
- Terminated Z-0347-2016
- Terminated Z-0350-2016
- Terminated Z-0348-2016
- Terminated Z-0351-2016
Monitor Ceiling Suspension system may fall
-
- Terminated Z-1280-2016
Loss of key image functionality due to a bent pedal of the Footswitch.
-
- Terminated Z-2976-2011
System freezes from 20 seconds up to one minute. No fluoroscopy exposure and no geometry movements during this time period.
-
- Terminated Z-1368-2011
Adjustment of the micro-switches in the footswitch used on the Philips cardiovascular X-ray systems was not implemented. If it is programmed for exposure, it is possible that after pressing the pedal on the footswitch it does not release itself. The patient may be exposed to X-ray longer than intended.
-
- Terminated Z-0124-2011
Due to a potential geometry software lock up during a Digital Rotational Angiography (DRA) run, the stand and table movement may no longer be possible.
-
- Terminated Z-0125-2011
Potential for table and stand not to move due to a software error.
-
- Terminated Z-1642-2008
Potential for radiologic patient table to become immobile and unable to move again due to force sensor sensitivity to electromagnetic radiation.
-
- Terminated Z-0162-2007
- Terminated Z-0161-2007
Potential for x-ray system to freeze during an examination, and inability of table to be re-positioned.
-
- Terminated Z-0208-06
- Terminated Z-0209-06
Potential for high voltage exposure involving the injector interface connector of patient support tables.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IZI
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1600
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent