Allura Xper FD20
by Philips Medical Systems North America, Inc.
510(k) K033737
Active
Valid until
No expiry
FDA recalls
15 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK033737 | Date9 Dec 2003 | StatusActive |
Recalls
-
- Ongoing Z-1148-2025
- Ongoing Z-1160-2025
- Ongoing Z-1151-2025
- Ongoing Z-1144-2025
- Ongoing Z-1158-2025
- Ongoing Z-1154-2025
- Ongoing Z-1149-2025
- Ongoing Z-1142-2025
- Ongoing Z-1150-2025
- Ongoing Z-1145-2025
- Ongoing Z-1143-2025
- Ongoing Z-1159-2025
- Ongoing Z-1147-2025
Possibility of the patient falling from the table related to the mattress used on the patient table.
-
- Ongoing Z-1080-2025
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
- Ongoing Z-1144-2024
Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
-
- Terminated Z-1140-2019
- Terminated Z-1141-2019
Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.
-
- Terminated Z-3223-2018
- Terminated Z-3221-2018
- Terminated Z-3222-2018
- Terminated Z-3224-2018
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
-
- Terminated Z-1822-2017
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
-
- Terminated Z-1062-2017
- Terminated Z-1061-2017
System may lock up.
-
- Terminated Z-0922-2017
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.
-
- Terminated Z-0764-2016
Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced position, the air gap of the brake can become too large, resulting in continued C-arc movement (C-arc is looking for balanced position).
-
- Terminated Z-0810-2016
The actuator assembly became detached and the monitor carriage with a flexvision monitor dropped to its lowest position.
-
- Terminated Z-0346-2016
- Terminated Z-0351-2016
- Terminated Z-0348-2016
- Terminated Z-0350-2016
- Terminated Z-0347-2016
- Terminated Z-0349-2016
- Terminated Z-0352-2016
- Terminated Z-0354-2016
- Terminated Z-0353-2016
Monitor Ceiling Suspension system may fall
-
- Terminated Z-1280-2016
Loss of key image functionality due to a bent pedal of the Footswitch.
-
- Terminated Z-0182-2015
- Terminated Z-0181-2015
Faulty Automatic Motion Controller (AMC), a problem in the Power On Self Test (POST) error handling was detected, can result in a hazardous movement of the C-arc. system.
-
- Terminated Z-1462-2012
The monitor Ceiling suspension may fail, which may cause the monitor to drop.
-
- Terminated Z-2976-2011
System freezes from 20 seconds up to one minute. No fluoroscopy exposure and no geometry movements during this time period.
-
- Terminated Z-1368-2011
Adjustment of the micro-switches in the footswitch used on the Philips cardiovascular X-ray systems was not implemented. If it is programmed for exposure, it is possible that after pressing the pedal on the footswitch it does not release itself. The patient may be exposed to X-ray longer than intended.
-
- Terminated Z-2448-2010
A lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem could only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted.
-
- Terminated Z-2449-2010
A lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem can only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted.
-
- Terminated Z-0349-2009
- Terminated Z-0348-2009
Tube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
-
- Terminated Z-2476-2010
- Terminated Z-2475-2010
A high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
-
- Terminated Z-0329-2008
- Terminated Z-0330-2008
Inaccurate anatomical measurements when using the optional SmartMask feature
-
- Terminated Z-0166-2007
Potential for generator to lock-up. Fluoroscopy and x-ray is no longer possible, and system must be restarted to be able to continue.
-
- Terminated Z-0582-06
Potential lock-up of system requiring reboot of system.
-
- Terminated Z-0209-06
- Terminated Z-0208-06
Potential for high voltage exposure involving the injector interface connector of patient support tables.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- OWB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent