Allura Xper or Table Series
by Philips Medical Systems North America, Inc.
510(k) K102005
Active
Valid until
No expiry
FDA recalls
36 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK102005 | Date9 Aug 2010 | StatusActive |
Recalls
-
- Ongoing Z-2893-2026
- Ongoing Z-2891-2026
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
- Ongoing Z-2412-2026
Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection
-
- Ongoing Z-2259-2026
- Ongoing Z-2258-2026
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
-
- Ongoing Z-1066-2026
- Ongoing Z-1073-2026
- Ongoing Z-1074-2026
- Ongoing Z-1064-2026
- Ongoing Z-1069-2026
- Ongoing Z-1077-2026
- Ongoing Z-1070-2026
- Ongoing Z-1067-2026
- Ongoing Z-1072-2026
- Ongoing Z-1076-2026
- Ongoing Z-1065-2026
- Ongoing Z-1068-2026
- Ongoing Z-1071-2026
- Ongoing Z-1062-2026
- Ongoing Z-1063-2026
- Ongoing Z-1075-2026
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
- Ongoing Z-2624-2025
- Ongoing Z-2627-2025
- Ongoing Z-2625-2025
- Ongoing Z-2631-2025
- Ongoing Z-2633-2025
- Ongoing Z-2634-2025
- Ongoing Z-2629-2025
- Ongoing Z-2630-2025
- Ongoing Z-2632-2025
- Ongoing Z-2635-2025
- Ongoing Z-2623-2025
- Ongoing Z-2628-2025
- Ongoing Z-2626-2025
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
- Ongoing Z-2504-2025
Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
-
- Ongoing Z-2503-2025
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
-
- Terminated Z-1822-2017
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
-
- Terminated Z-0765-2016
Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced position, the air gap of the brake can become too large, resulting in continued C-arc movement (C-arc is looking for balanced position).
-
- Terminated Z-1120-2015
In certain circumstances, a software error can lead to a situation where the five minute fluoroscopy audible signal does not sound.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- OWB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent