Alm X'ten (X10) Surgical Light System
by Alm S.A.
510(k) K040735
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK040735 | Date21 May 2004 | StatusActive |
Recalls
-
- Ongoing Z-0766-2024
A potential for a light system to fall in the operating room.
-
- Terminated Z-1927-2018
Getinge has received complaints concerning the X TEN Surgical Lights whose spring arms have broken, causing light heads to separate from the assembly. The issue relates to cracking of the spring arms at the weld seams that has the potential to develop into breaks with heavy use and aging of the equipment.
-
- Terminated Z-1990-2015
Maquet Suspension Arms used with PowerLED/X-TEN/Volista Surgical Light Systems may grind at the junction of the suspension arms which can cause paint chips or particles to fall into the surgical field, when the suspension arms are mounted above the surgical table.
-
- Terminated Z-0186-2010
- Terminated Z-0187-2010
- Terminated Z-0188-2010
Several reports have been received regarding one model of spring arm (Acrobat 200 series) in combination with surgical light from various Manufacturers. All incidents pertain to a crack on the front joint of the Acrobat arm.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- FSY
- Device Class
- Class II
- Regulation
- 21 CFR Part 878.4580
- Advisory Committee
- General, Plastic Surgery
- FDA Decision
- SESE Substantially Equivalent