Amicas Pacs, Version 6.0

by Amicas, Inc.

510(k) K082144

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

13 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK082144 Date5 Sep 2008ClearedStatusActive

Recalls

13 recalls · FDA

  • 2 recalls, same reason

    • Terminated Z-2601-2017 FDA Class II initiated 24 May 2017
    • Terminated Z-1431-2017 FDA Class II initiated 4 Apr 2016

    Studies that have QC, PDE, or HL7 ADT updates applied may fail to propagate via telmed protocol to the destination server and may fail to import on the destination server.

    • Terminated Z-2590-2017 FDA Class II initiated 8 May 2017

    Merge PACS did not show unviewed images when the last view was skipped. There is a potential risk to health of a physician misdiagnosis because not all images are available for viewing.

    • Terminated Z-2591-2017 FDA Class II initiated 8 May 2017

    When cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis.

    • Terminated Z-1176-2017 FDA Class II initiated 4 Apr 2016

    Potential exists for an incorrect patient image being displayed which could result in the delay in diagnosis or treatment.

    • Terminated Z-0599-2017 FDA Class II initiated 28 Mar 2016

    The software did not show unviewed images when the last view was skipped. The physician may potentially misdiagnose when not all images are available for viewing.

    • Terminated Z-0726-2017 FDA Class II initiated 28 Mar 2016

    Cut lines on the image may present horizontally rather than vertically.

    • Terminated Z-2363-2016 FDA Class II initiated 5 Feb 2016

    When measuring a lesion on an unmagnified mammography image then performing the same measurement on an image magnified by the Hologic imager (different image), the measurements are not the same.

  • 2 recalls, same reason

    • Terminated Z-2532-2016 FDA Class II initiated 30 Jan 2016
    • Terminated Z-2531-2016 FDA Class II initiated 30 Jan 2016

    Studies that are viewed in iConnect Access that originate on Merge PACS / Ortho PACS may not display the correct current patient demographics if there has been a PDE (Patient / Study Demographic Edit) performed after the images were ingested in PACS.

    • Terminated Z-2301-2016 FDA Class II initiated 30 Jan 2016

    Studies coming over via telmed were missing patient's DOB, procedure, and referring physician.

    • Terminated Z-2303-2016 FDA Class II initiated 30 Jan 2016

    A migrated study that has annotations will display the annotations, but when a prior study is loaded that should have annotations, the second study will not display annotations.

    • Terminated Z-2379-2016 FDA Class II initiated 30 Jan 2016

    The patient name in the Halo title bar and the thumbnails do not match the name on displayed images.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LLZ
Device Class
Class II
Regulation
21 CFR Part 892.2050
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent