Ammonia 05975581190
Ammonia 05975581190 is an ammonia IVD (FDA Class I; FDA product code JIF, Enzymatic Method, Ammonia) made by Roche Diagnostics GmbH, cleared under FDA 510(k) K972250 on 12 Aug 1997 and listed in FDA GUDID since 8 Feb 2021 (UDI-DI 04015630928071), active as of 12 Nov 2026.
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number04015630928071 | Date8 Feb 2021 | StatusListed |
| United States | FDA 510(k)SESE | NumberK972250 | Date12 Aug 1997 | StatusActive |
Official records
- AccessGUDID record 04015630928071
- FDA 510(k) database K972250
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K972250 decided 12 Aug 1997 Active
- FDA GUDID UDI-DI 04015630928071 listed 8 Feb 2021 Listed
Recalls
-
- Terminated Z-0688-2008
False Results- The reagent of the LDL-Cholesterol assay shows a carryover effect on the Lipase assay. Results may be falsely elevated by up to 50 units per liter when used on COBAS INTEGRA 400 or COBAS INTEGRA 400+ analyzers.
Timeline
- Cleared FDA 510(k) K972250
- Recall initiated FDA recalls Z-0688-2008
- Listed FDA GUDID 04015630928071
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Ammonia
- Generic name
- Ammonia IVD, kit, spectrophotometry
- Model / Ref / Catalog
- 05975581190
- Primary DI / UDI-DI
- 04015630928071
- 510(k) Number
- FDA product code
- JIF
- GMDN term
- Ammonia IVD, kit, spectrophotometry
- Device class
- Class I
- Regulation number
- 862.1065
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code JIF (ENZYMATIC METHOD, AMMONIA).
- EASY-TEST AMMONIA (AMON) ITEM NO. 16695 by Em Diagnostic Systems, Inc.
- DIMENSION VISTA AMMONIA (AMON) FLEX REAGENT CARTRIDGE by Dade Behring, Inc.
- ACA AMMONIA ANALYTICAL TEST PACK by E.I. Dupont DE Nemours & Co., Inc.
- SYNCHRON SYSTEMS AMMONIA REAGENT by Beckman Coulter, Inc.
- BECKMAN AMMONIA AMM REAGENT/ CX AMMONIA CALIBRATOR by Beckman Instruments, Inc.
- AMMONIA (AMON) TESTPACK, ITEM NO. 67689/95 by Em Diagnostic Systems, Inc.
- AMMONIA REAGENT by Gilford Diagnostics
- AMMONIA-INCORPORATING DYNAMIC STABILIZATION TECHNOLOGY (DST) by Trace America, Inc.
- AMMONIA ASSAY FOR THE ADVIA INTEGRATED MODULAR SYSTEM by Bayer Diagnostics Corp.
- SYNCHRON SYSTEMS AMMONIA (AMM) REAGENT by Beckman Coulter, Inc.