Ana Hep-2 Fluoro-Kit

by Clinical Sciences, Inc.

510(k) K822397

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 11 Oct 2026. What this means

Regulatory status

as of registry sync 11 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK822397 Date21 Sep 1982ClearedStatusActive

Recalls

2 recalls · FDA

    • Terminated Z-0807-05 FDA Unclassified (pre-2012) initiated 24 Sep 2004

    A problem with software for the PolyTiter Immunofluorescent Titration System in that no flags or warnings appear when PolyTiter Calibrator Values are re-entered or changed (including values changed to unacceptable values) after the initial curve has been generated, if the 'refresh button' has not been pressed. The graph may appear valid; however erroneous results may occur.

    • Terminated Z-0676-05 FDA Unclassified (pre-2012) initiated 30 Jun 2004

    The program for a specific immunofluorescent titration system (Polymedco PolyTiter immunofluorescent Titration System) was found faulty in that the calibrator 2 value was incorrectly programmed as 1:120 instead of 1:160.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
DHN
Device Class
Class II
Regulation
21 CFR Part 866.5100
Advisory Committee
Immunology
FDA Decision
SESE Substantially Equivalent

Other devices from Clinical Sciences, Inc.

See all devices from Clinical Sciences, Inc. →