Anatomical Shoulder System/Anatomical Shoulder Domelock System/Anatomical Shoulder Fracture System
by Zimmer GmbH
510(k) K142403
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK142403 | Date24 Nov 2014 | StatusActive |
Recalls
-
- Terminated Z-2003-2016
In some cases it has been difficult or not possible to disassemble the adjusted Dome Centric from the AS Humeral Rasp after initial positioning of the AS Humeral Trial Head which resulted in the whole construct of the Dome Centric with the still assembled AS Humeral Rasp taken out of the humeral canal. This could cause a delay in surgery time or the surgeon could decide to close the wound without finishing the surgery, increasing the infection risk or second use of anesthesia necessary.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- KWS
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3660
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent