AneurysmFlow
by Philips Medical Systems Nederland B.V. Netherlands EUDAMED510(k)GUDID
Ref 001015510(k) K160455
Product code OWBEMDN Z11030192
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number0884838BM321SH | Date17 Oct 2024 | StatusActive |
| United States | GUDID listing GUDID | Number00884838064768 | Date3 Mar 2017 | StatusIn distribution |
| United States | FDA 510(k)SESE 510(k) | NumberK160455 | Date25 Aug 2016 | StatusActive |
Coverage
- Active
- Not covered
- United States
- European Union
Recalls
-
- Ongoing Z-1805-2026
Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.
Specification
- Generic Name
- Cardiovascular fluoroscopy/intravascular imaging software
- Model / Ref / Catalog
- 001015, AneurysmFlow - 1
- Primary DI / UDI-DI
- 00884838064768
- Basic UDI-DI
- 0884838BM321SH
- 510(k) Number
- FDA Product Code
- OWB, LLZ
- EMDN Code
- Z11030192 Digital angiography systems - medical device software
- GMDN Term
- Cardiovascular fluoroscopy/intravascular imaging software
- Device Class
- Class IIb EUDAMEDClass II GUDID
- Regulation Number
- 892.1650, 892.2050
- Special Device Type
- Software
- Market of Registration
- Netherlands
- Commercial Distribution
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / Reference | 001015 | AneurysmFlow - 1 |
| Device Class | Class IIb | Class II |
Manufacturer details
- EUDAMED SRN
- NL-MF-000001489
Other devices from Philips Medical Systems Nederland B.V.
See all devices from Philips Medical Systems Nederland B.V. →
Properties & raw data
Show properties, registry sources and raw records
Other properties (21)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No
Sources (3)
- EUDAMED fb4ff8aa-3267-4280-b3f1-581676df3acc
- FDA UDI 9ecaff46-3783-469e-a39b-f56a41b67e85
- FDA 510(k) K160455